Skip to main content

Table 3 The Maximum von Mises stress values in framework (MPa)

From: Comparative biomechanics of all-on-4 and vertical implant placement in asymmetrical mandibular: a finite element study

Model

Anterior-Vertical loading (100 N)

Posterior-Vertical

loading (200 N)

Posterior-Oblique

loading (200 N)

All-on-4

33.1

89.3

213.8

All-on-5-o

61.2

40.2

161.5

All-on-5-v

37.6

58.7

228.9