Skip to main content

Table 2 Calibration and sensitivity data of monomers

From: Effect of topical fluoride applications on residual monomer release from resin-based restorative materials

 

Regression equation

R2

RT

% RSD

LOD (μg/ml)

LOQ (μg/ml)

HEMA

y = 36.398x + 6.287

0.99997

4.991

7.313

0.663

2.010

TEGDMA

y = 43.491x + 5.720

0.99999

3.150

4.903

0.372

1.127

UDMA

y = 17.161x + 3.469

0.99996

6.565

3.840

0.738

2.237

Bis-GMA

y = 35.240x + 2.079

0.99994

7.990

9.042

0.846

2.565